Beer-Lambert law calculator
A = ε · l · c
Fill in three of the four fields and leave the one you need empty. An absorbance above 2.0 is flagged: on a routine spectrophotometer stray light makes such readings too low.
Result
Enter any 3 values
About this calculation
Runs in your browser
The calculation runs in your browser. Your inputs are kept in this browser's storage on this device so they are still here when you come back; Clear removes them.
Checked
Checked against: the hand-worked NADH example, solved for each of the four quantities.
Calculation version 088929cdc6c2, shared with the Android app.
Worked example
NADH at 340 nm (ε = 6220 L·mol⁻¹·cm⁻¹) in a 1 cm cuvette reads A = 0.622.
c = 0.622 / (6220 × 1) = 1.000 × 10⁻⁴ mol/L
The absorbance is well below 2, so the reading is inside the range a routine spectrophotometer measures linearly.
The Load example button fills in these values.
Formula
A = ε · l · c
- A
- absorbance, dimensionless (A = −log₁₀ T)
- ε
- molar (decadic) absorption coefficient, L·mol⁻¹·cm⁻¹
- l
- optical path length, cm
- c
- amount concentration, mol/L
Leave one field empty and it is calculated from the other three. The units are fixed: convert a path length in mm or a concentration in mmol/L before entering it.
Assumptions and limits
- ε belongs to one substance at one wavelength, in a given solvent and at a given pH and temperature. Use a value determined under your conditions.
- The law holds for dilute solutions and monochromatic light. Above about A = 2, stray light and the finite bandwidth make the measured absorbance read low, so the concentration is underestimated. The calculator flags A above 2.0.
- Only the analyte absorbs: the blank is subtracted, and scattering, fluorescence and other absorbing species are absent.
- For quantitative work a calibration curve measured on your instrument is more reliable than a literature ε.
- Results are shown to six significant figures; measurement uncertainty is not included.
References
The relation is the Beer-Lambert law itself; it uses no tabulated data. The example uses ε = 6220 L·mol⁻¹·cm⁻¹ for NADH at 340 nm, the value biochemical assays are commonly standardised on. For your own work, use an ε determined under your conditions.
The calculation is the one used by the Lambert-Beer Law calculator in the LabTools Android app.