HPLC column volume and flush calculator
V = π · (d/2)² · L / 1000
Enter the column length and internal diameter, the flow rate and the number of column volumes to flush. The volume, the flush volume and the time update as you type.
Results
Enter the column length and internal diameter
About this calculation
Runs in your browser
The calculation runs in your browser. Your inputs are kept in this browser's storage on this device so they are still here when you come back; Clear removes them.
Checked
Checked against: a hand-worked 250 × 4.6 mm column, the flush volume and time, and the limits of the flush factor.
Calculation version 088929cdc6c2, shared with the Android app.
Worked example
A 250 mm × 4.6 mm column flushed with 20 column volumes at 1 mL/min.
V = π · (4.6 / 2)² · 250 / 1000 = 4.155 mL; flush = 20 × 4.155 = 83.095 mL; time = 83.095 / 1 = 83.1 min (1 h 23 min)
The Load example button fills in these values.
Formula
V = π · (d / 2)² · L / 1000
- L
- column length, mm
- d
- internal diameter, mm
- V
- empty (geometric) column volume, mL
flush volume = flush factor × V; flush time = flush volume / flow rate
Assumptions and limits
- This is the empty, geometric volume of a cylinder. The volume of mobile phase inside a packed column is smaller, because the packing takes up part of it; the app's Dead Volume & Kinetics tab, which uses the column porosity, is part of PRO.
- The tubing, the fittings, the injector and the detector cell are not included.
- The flush factor is a number of column volumes (CV), from 1 to 100. The app's guidance is 5–10 CV to equilibrate with a mobile phase or buffer, 10–20 CV to condition a new column or exchange the solvent, and 30–100 CV to clean and regenerate. Use the figures in your column's care instructions where they differ.
- The time needs the flow rate; without it the volumes are given alone. The time takes the flow as constant and does not check it against the column's pressure limit.
References
The volume is the geometry of a cylinder, and the flush time is a volume divided by a flow rate: no tabulated data is used. The flush guidance is that recorded in the LabTools Android app.
The calculation is the one used by the Column & Dead Volume (LC) calculator in the app.