Mobile phase mixer
V = V_total × %
Enter the total volume, the solvents with their percentages and, if you need them, additives. The volume or weight to measure out for each bottle appears as you type, with a tick-off list for the bench.
Volumes to measure out
Enter at least 1 solvent and a total volume to calculate
About this calculation
Runs in your browser
The calculation runs in your browser. Your inputs are kept in this browser's storage on this device so they are still here when you come back; Clear removes them.
Checked
Checked against: a hand-worked acetonitrile and water blend with a formic acid additive in volume and weight, a solid and a liquid additive, on-line mixing and the warnings.
Calculation version 088929cdc6c2, shared with the Android app.
Worked example
1000 mL of a mobile phase of acetonitrile and water 60:40 with 0.1 % formic acid.
formic acid 0.1 % × 1000 mL = 1 mL; remainder 999 mL; acetonitrile 0.60 × 999 = 599.4 mL; water 0.40 × 999 = 399.6 mL
In weight mode, with densities of 0.786 g/mL for acetonitrile and 0.998 g/mL for water at 20 °C, that is 471.1 g of acetonitrile and 398.8 g of water. The Load example button fills in these values.
Formula
Vcomponent = Vtotal × percentage
Vadditive = Vtotal × % ; Vsolvent i = (Vtotal − ΣVadditive) × % solvent i
Additives are subtracted from the total volume before the solvents split what remains, so the final volume is the one you entered. A liquid additive is a dilution of a stock (C₁V₁ = C₂V₂, against the same total volume); a solid additive is a mass, m = C · M · V, from a molar concentration, a molar mass and the total volume. In weight mode the volume of each recognised liquid is multiplied by its density.
Assumptions and limits
- Solvent percentages (% v/v) should add up to 100 %. A sum other than 100 % is flagged: within 5 % as a warning, further off as an error.
- Volumes are treated as additive. For water with a miscible organic solvent the mixed volume is measurably smaller than the sum of the parts: a published HILIC eluent article puts the loss at about 20 mL when 500 mL of water and 500 mL of acetonitrile are mixed. For a method written as volume ratios, measure each component separately and then mix them; making up to the mark with one solvent gives more of that solvent than intended. Document the way you prepare it.
- Adjust the pH of the aqueous or buffer phase before adding the organic modifier: a pH electrode reads incorrectly in the presence of organic solvent.
- On-line mixing (shown with two or more solvents): a gradient pump blends the solvent lines from the bottles on the instrument, so percentage additives are folded into the water bottle, or the largest one, instead of being measured on their own.
- A solid additive displaces volume, so the largest solvent is shown as "fill up to" the final mark instead of a measured volume.
- Weight mode uses the densities of the solvents the calculator recognises by name (exact match of the name or a common abbreviation, such as ACN or MeOH), at about 20 °C. Ammonium hydroxide is taken as the common 28–30 % reagent, not neat ammonia. It is not shown for liquid additives made from a stock, or for names it does not recognise.
- Densities of water–organic mixtures differ from those of the pure liquids; the weights are those of the separate components.
References
The relations are definitions: a percentage by volume of a total, the dilution equation, and mass from molar concentration. The densities are those recorded in the LabTools Android app, which cites the CRC Handbook of Chemistry and Physics for them.
On the volume contraction and on preparing mixtures: Eluent preparation for hydrophilic interaction liquid chromatography, part 1: solvent volumes and buffer concentration (LCGC), and the Merck Millipore technical article HPLC tips & tricks: mobile phase preparation.
The calculation is the one used by the Mobile Phase Mixer in the LabTools Android app.