TDS calculator for ICP-MS and ICP-OES
TDS vial = TDS · aliquot / final volume
Choose a solid or a liquid sample, give its mass or its TDS, the aliquot and the final volume, and the instrument and mode. The TDS in the vial, a verdict and the recommended dilution appear.
Result
Enter sample and preparation parameters
About this calculation
Runs in your browser
The calculation runs in your browser. Your inputs are kept in this browser's storage on this device so they are still here when you come back; Clear removes them.
Checked
Checked against: a hand-worked digested solid and a diluted vial against the limits of ICP-MS and ICP-OES, each instrument mode, the conductivity route and the thresholds.
Calculation version 088929cdc6c2, shared with the Android app.
Worked example
0.5 g of a solid is digested and made up to 50 mL, and 1 mL of the digest is diluted to 10 mL for an ICP-MS.
TDS of the digest = 0.5 g / 50 mL = 10 000 mg/L; dilution factor 10; in the vial 1000 mg/L = 0.1 %
That is below the 0.2 % limit used for a standard ICP-MS, so the sample is safe. The Load example button fills in these values.
Formula
TDSvial = TDSinitial × aliquot / final volume
For a digested solid, the initial TDS is the sample mass divided by the volume it was made up to (1 g in 100 mL is 10 000 mg/L, or 1 %); for a liquid it is a known value, a conductivity reading × 0.65, or a typical value for the kind of sample. The result in % is mg/L ÷ 10 000.
With the AGD mode of an ICP-MS, additional argon is mixed into the aerosol and dilutes the matrix before it reaches the plasma; the calculator gives the aerosol dilution factor that brings the vial down to 0.2 %, rounded up.
Assumptions and limits
- The limits are rules of thumb, not properties of every instrument. A standard ICP-MS: 0.2 % TDS (2 g/L), a figure also written into standard methods for ICP-MS, because undissociated matrix deposits on the interface cones and causes drift. A standard ICP-OES nebulizer: a warning at 2 % and critical above 3 %. High-solids nebulizers: 15 % and 30 %. Manufacturers publish their own ratings for a given nebulizer or accessory, and they differ: use the data sheet of your instrument.
- For a solid, the whole sample is taken as dissolved in the digest. That is the worst case: a sample that does not dissolve completely, or that loses mass (carbon, silicon) in the digestion, carries less.
- TDS from conductivity is TDS ≈ 0.65 × EC (µS/cm). It is an empirical correlation (APHA Standard Methods 1030 E, which gives 0.55–0.70 for potable waters), and the factor moves with the ionic composition, so use a known or measured TDS whenever one is at hand.
- The templates are rough, typical TDS values of a kind of sample (drinking water about 500 mg/L, seawater about 35 000 mg/L, and so on) to size a first dilution. They are not values to report.
- TDS says nothing about which elements are present, and it does not replace checking an internal standard, signal drift and the cones during a run.
References
For ICP-MS, the limit of 0.2 % (2 g/L) TDS is the one stated in Agilent's ICP-MS application notes and in standard methods such as EN ISO 17294-2 and US EPA 6020.
The conductivity correlation is that of APHA Standard Methods for the Examination of Water and Wastewater, section 1030 E (checking the correctness of analyses; the 0.55–0.70 range is stated there for potable waters). The thresholds and templates are those recorded in the LabTools Android app.
The calculation is the one used by the TDS (ICP) calculator in the app, where PDF, Excel and CSV export is part of PRO.