Brix ↔ refractive index
°Bx at 20 °C ↔ nD
Enter what the refractometer showed and the temperature of the sample. The reading at 20 °C, in °Bx and as a refractive index, appears with the temperature correction.
Result
Enter what the instrument showed and the temperature it showed it at.
The scale is defined for pure sucrose in water. For juice, wine must, honey or a coolant, read the result as apparent sucrose (°Bx equivalent), not as total solids.
About this calculation
Runs in your browser
The calculation runs in your browser. Your inputs are kept in this browser's storage on this device so they are still here when you come back; Clear removes them.
Checked
Checked against: the refractive index of a 20 °Bx sucrose solution, the temperature correction at 25 °C and the limits of the published tables.
Calculation version 088929cdc6c2, shared with the Android app.
Worked example
A refractometer without temperature compensation reads 20.0 °Bx on a sample at 25 °C.
20.38 °Bx at 20 °C (correction +0.38 °Bx), nD = 1.36447
The warm sample read low, so the correction is added. The Load example button fills in these values.
Formula
°Bx is grams of sucrose per 100 g of solution. A refractometer measures the refractive index and reads °Bx off a scale that converts one to the other. That scale is defined at 20 °C.
°Bx at 20 °C = reading + correction(reading, T)
Two reference tables are used: the refractive index of sucrose solutions at 20 °C on the sodium D line, held at every whole percent from 0 to 95 °Bx and interpolated between, and the published refractometric temperature correction, held at 1 °C by 5 °Bx over 15–40 °C and 0–85 °Bx and interpolated in both directions. A refractive index entered instead is turned into its apparent °Bx first, so both routes give the same answer.
Assumptions and limits
- The scale is defined for pure sucrose in water. For juice, wine must, honey or a coolant, read the result as apparent sucrose (°Bx equivalent), not as total solids.
- Leave the temperature empty when the instrument has automatic temperature compensation: its reading is already at 20 °C.
- The published correction table covers 15–40 °C and 0–85 °Bx. Outside that it ends, and the value shown is held at the nearest edge rather than extrapolated — treat it as indicative only.
- Off the scale — Brix runs 0 to 95, refractive index 1.33299 to 1.53200.
- The correction is looked up at the reading, which is how the published table is indexed and how a correction card is used at the bench.
References
The refractive index scale is the ICUMSA/NBS reference scale for sucrose solutions at 20 °C, which defines °Bx. The temperature correction is the published refractometric correction table, transcribed into the calculation code with its limits (15–40 °C, 0–85 °Bx) kept.
The calculation is the one used by the Brix ↔ Refractive Index converter in the LabTools Android app.