Radiation dose converter
H = D · wR
Choose the quantity you have, the unit you need and, when you cross from absorbed to equivalent dose, the type of radiation.
Result
Enter a dose
About this calculation
Runs in your browser
The calculation runs in your browser. Your inputs are kept in this browser's storage on this device so they are still here when you come back; Clear removes them.
Checked
Checked against: the ICRP 103 weighting factors, the exact rad and rem definitions and a hand-worked dose over time.
Calculation version 088929cdc6c2, shared with the Android app.
Worked example
An absorbed dose of 2 mGy:
photons: H = 2 mGy × 1 = 2 mSv; alpha particles: H = 2 mGy × 20 = 40 mSv
A survey meter reading 0.5 µSv/h over 3 hours accumulates 1.5 µSv, which is 0.00015 rem. The dose rate in mrem/h for 0.5 µSv/h is 0.05. The Load example button fills in these values.
Formula
H (Sv) = D (Gy) × wR
- D
- absorbed dose in gray: the energy absorbed per kilogram
- H
- equivalent dose in sievert: that energy weighted by how damaging the kind of radiation is
- wR
- radiation weighting factor
wR (ICRP Publication 103, 2007): photons 1, electrons 1, protons 2, alpha particles and heavy ions 20. Within one quantity (Gy ↔ rad, Sv ↔ rem, or a prefix) no weighting is used: 1 Gy = 100 rad and 1 Sv = 100 rem exactly. A dose rate converts between per-hour and per-second units by a plain factor. Dose over time multiplies a constant dose rate by the exposure time.
Assumptions and limits
- This is equivalent dose to a tissue, not effective dose, which also weights by organ. No dose limits are shown: they depend on country, occupation and year.
- For X-ray fluorescence analysers, X-ray diffraction and gamma sources the radiation is photons, so Gy and Sv are numerically equal.
- Neutrons are not offered: their weighting factor is a continuous function of energy.
- A work year is taken as 2000 h (50 weeks of 40 h). A calendar year is deliberately not offered, because it is 8760 or 8766 h depending on the source.
- The dose rate is taken as constant over the exposure time.
References
The radiation weighting factors are those of ICRP Publication 103 (2007), Table 4.2. The rad (0.01 Gy) and the rem (0.01 Sv) are defined exactly.
The calculation is the one used by the Radiation dose and dose rate converter in the LabTools Android app.