SLPM ↔ ALPM converter

Qa = Qs · (Ps/Pa) · (Ta/Ts)

Enter the flow, the actual temperature and pressure, say whether the pressure is absolute or gauge, and choose the reference condition of the standard flow.

Result

Enter a flow, a temperature and a pressure

About this calculation

Runs in your browser

The calculation runs in your browser. Your inputs are kept in this browser's storage on this device so they are still here when you come back; Clear removes them.

Checked

Checked against: a hand-worked conversion at 25 °C and 1.5 bar, the reference conditions and the gauge to absolute step.

Calculation version 088929cdc6c2, shared with the Android app.

Worked example

A mass-flow controller delivers 10 standard L/min (referenced to 0 °C and 101.325 kPa). The gas is at 25 °C and 1.5 bar absolute.

Qa = 10 × (101.325 / 150) × (298.15 / 273.15) = 7.373 L/min

The Load example button fills in these values.

Formula

Qactual = Qstandard · (Pstd / Pact) · (Tact / Tstd)

Qstandard = Qactual · (Pact / Pstd) · (Tstd / Tact)

Temperatures are absolute (kelvin). The flow unit cancels, so the same ratio serves SLPM ↔ ALPM, sccm ↔ accm and Nm³/h ↔ m³/h.

Assumptions and limits

  • A standard flow is not a volume per time. A mass-flow controller measures mass and reports it as the volume that mass would occupy at a stated reference temperature and pressure. SLPM and ALPM are the same throughput described twice, and moving between them is the ideal-gas law with the molar flow held fixed.
  • There is no single standard. ISO 13443 defines 0 °C as the "normal" and 15 °C as the "standard" condition (both at 101.325 kPa); 20 °C and 25 °C are also in common use. Manufacturers differ: many mass-flow controllers are referenced to 0 °C, but at least one manufacturer's data sheets give 25 °C as the default, and the reference is usually configurable. 0 °C against 25 °C is 9.2 %, and nothing in a number written "SLPM" says which was meant, so it is chosen here. Read the reference from the instrument or its certificate.
  • A gauge reads zero at ambient; the gas law does not care what ambient is. Entering 2 bar from a gauge as though it were absolute makes the result 51 % wrong, and the wrong number looks reasonable. Choosing Gauge adds one standard atmosphere, which is right in a sea-level laboratory and wrong by the altitude anywhere else, so the absolute pressure used is printed with the result and Absolute or Gauge has to be chosen before a result appears.
  • Ideal gas. For nitrogen, helium, argon, air and hydrogen below about 10 bar the compressibility correction is a few parts per thousand and inside a controller's own accuracy. For CO₂, ammonia or propane near their condensation conditions it is not: the calculator notes it above 10 bar absolute.

References

The reference conditions offered are 0 °C and 15 °C (the "normal" and the "standard" condition of ISO 13443), 20 °C and 25 °C, each at 101.325 kPa. That manufacturers use different references is stated in the data sheets of Alicat (standard 25 °C, normal 0 °C, both configurable).

The calculation is the one used by the SLPM ↔ ALPM converter in the LabTools Android app.